PATIENT MONITOR
MODEL M50
The new generation of EDAN patient monitors M50 provides perfect transportation solution. The multi-functional all-in-one parameter module and new application platform greatly enhance the efficiency of full clinical parameter monitoring for patient transfer.
Features:
- 8.4″ color TFT display with maximum 11 waveforms
- Pacemaker detection, electrosurgical interference proof
- Defibrillation protection and defibrillation synchronization
- Full touchscreen enables intuitive operation by clicking on the specific parameters or the waveform in real time, standard configuration includes the navigation knob
- OxyCRG available to judge the respiration and circulation function for neonates
- Large font display
- Arrhythmia and S-T segment analysis
- Built-in rechargeable Li-ion battery
- Nurse Call function and bi-directional communication with MFM-CMS central station
- SD card slot enables memory extension for large data storage
- USB and serial ports enables abundant future upgrades
- VGA output and analog output
- Complete and flexible mounting solution for different scenarios
Model |
Standard
|
Optional
|
M50 | 3/5-lead ECG, EDAN SpO2, RESP, NIBP, 2-Temp, PR | Touchscreen, Thermal Recorder, WLAN accessory, Nellcor-SpO2, 2-IBP, Respironics CO2 (for intubation and non-intubation application), 12V DC input |
Specification :
Safety Standards
CE Marking in accordance to CouncilDirective 2007/47/EC concerning Medical Devices IEC 60601-1+A1+A2;IEC 60601-1-2+A1;
Physical Specification
Device dimension: 260 mm (L) x 140 mm (W) x 205 mm (H)Weight: Standard Configuration 3.6 kg
Display
Color TFT LCD 8.4 Touchscreen OptionalResolution: 800 X 600 dpi Traces Displayed: up to 8 Waveforms Displayed: up to 11 Various Working Interface Selectable: Standard Monitoring Display Large Font Intensive Care Display Trend Graph/ Monitoring Co-Display Bed to Bed view Display (Optional) OxyCRG Dynamic View Display Drug Dose Calculation Interface Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Operation Environment
Ambient Temperature:-20 - 55 (-4 -131 F) Humidity: 15%-95% non-condensing
Power Supply
External Power Supply100-240V AC, 50/60HZ Pmax=80VA Internal Battery Power Supply Type: Rechargeable Li-ion Voltage: 14.8 V DC Capacitance: 2100 mAh (4200 mAh optional) Battery working period: up to 6 hours Recharging time < 150 minutes (2100mAh) < 360 minutes (4200mAh))
Resp
Method: Trans-thoracic impedanceOperation mode: Auto/ Manual RR Measurement range: Adult: 0~120 rPM Neonate/Pediatric: 0~150 rPM Resolution: 1rPM Apnea alarm threshold: 10s,15s,20s (default),25s,30s,35s,40s Alarm: Yes, audible and visual alarm, alarm events recallable Band width: 0.2-2.5Hz (-3dB) Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG
Lead type: 5-lead and 3-lead selectable3 leadwire cable: RA; LA; LL or R; L; F 5 leadwire cable: RA; LA; RL; LL; V or R; L; N; F; C Lead selection: 3-lead: I; II; III; 5-lead: I; II; III; aVR; aVL; aVF; V Gain selection: x0.125;x0.25; x0.5; x1; x2; x4; auto Sweep speed: 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s ECG HR Range: Adult/Pediatric: 15-300bpm; Neonate: 15-350bpm Resolution & accuracy: 1bpm or 1%, whichever is greater Filter: Diagnostic mode: 0.05-150Hz Monitoring mode: 0.5-40 Hz Surgical mode: 1-20Hz Protection: Withstand 5000VAC/50Hz voltage in isolation against Defibrillation and electrosurgical interference ST-Segment Detection: Measurement range: -2.0 mV~2.0mV Alarm range: -2.0 mV~2.0mV ST-Segment Arrhythmia analysis and catergorization : Yes ARR Alarm: Audible and visual alarm, alarm events recallable Pace maker detection: Yes, and 5 types abnormal status detectable IEC 60601-2-25; AAMI EC 11/EC 13 IEC 60601-2-27
NIBP
Method: Automatic OscillometricOperation modes: Manual/Automatic/Continuous Auto measurement time interval: Adjustable 1/2/3/4/5/10/15/30/60/90/120/240/480 Minutes Measurement unit: mmHg/kPa selectable Measurement types: Systolic, Diastolic, Mean Pressure range for Adults: Systolic: 40 - 270 mmHg Diastolic: 10 - 215 mmHg Mean: 20 - 235 mmHg Pressure range for Pediatrics: Systolic: 40 - 200 mmHg Diastolic: 10 - 150 mmHg Mean: 20-165 mmHg Pressure range for Neonates: Systolic: 40 - 135 mmHg Diastolic: 10 - 100 mmHg Mean: 20 - 110 mmHg Leak test and pressure auto calibration: Yes Over-pressure protection: Dual Safety protection Resolution: 1mmHg Accuracy: Max mean error 5 mmHg Max standard deviation 8 mmHg Alarm: Systolic, Diastolic, Mean PR from NIBP: Measurement 40~240 bpm Resolution 1 bpm Accuracy 3bpm or 3% whichever is greater Leak test and pressure auto calibration: Yes IEC 60601-2-30
EDAN SpO2
Measurement & Alarm Range: 0 - 100%Resolution: 1%; Accuracy: 2% (70-100%, Adult/Pediatric); 3% (70-100%, Neonate) PR Measurement and Alarm Range: 30 - 300bpm Resoluton: 1 bpm; Accuracy: 3bpm Refresh rate: 2s ISO 9919
Nellcor SpO2 (optional)
Measurement & Alarm Range: 0 - 100%Resolution: 1%; Accuracy: 2-3% (70-100%, Adult/Pediatric); 3-3.5% (70-100%, Neonate) PR Measurement and Alarm Range: 20 - 300 bpm Resoluton: 1 bpm; Accuracy: 3bpm (depends on probe) Refresh Rate: 1s ISO 9919
Temperature (2 Channels, 1 probe by default)
Measurement /Alarm range: 0~50 (32-122 F)Probe type: YSI (B Series) and CF-FI Resolution: 0.1 Accuracy: 0.1 (without probe) Channel: Dual-channel. Provide T1; T2; T IEC 12470-4
Quick Temperature (Optional)
Measurement /Alarm range: 25~45Probe type: Oral/ Axillary / Rectal Resolution: 0.1 Accuracy: 0.1 (without probe) Typical measurement time <15s IEC 12470-4
IBP (2-Channels, optional)
Measuremed Pressure:ART, PA, CVP, RAP, LAP, ICP, P1,P2 Measurement range: -50 - 300 mmHg; Resolution: 1 mmHg Accuracy: 2% or 1 mmHg, whichever is greater (without probe) Sensitivity: 5 ?V/V/mmHg;Impedance range:300-3000 IEC 60601-2-34
CO2 (Mainstream and Sidestream, optional)
By Philips Respironics CAPNOSTAT 5 & LoFlo TechnologyRange: 0~ 150mmHg Accuracy: 2% 0 ~ 40mmHg, 5% 41~70mmHg 8% 71~100mmHg 10% 101~150 mmHg AwRR Accuracy: 1 rpm Convenient design for intubbated and non-intubated applciations Possible to work at low sample flow rate: 50ml / minute Detailed specification refer to the user manual of Respronics ISO 21647
Recorder (optional)
Built-in, direct thermal pixel array recorderUp to 3 channels printing 1,2, 3 channels selectable Print speed: 25mm/s, 50mm/s Paper width: 50 mm/Record width:48mm
I/O Interface)
2-USB PortsSD Card Socket RJ-45 Ethernet Port. IEEE 802.3 VGA Output Analog & Nurse Call output Defibrillation Synchronization output WLAN Access Point 802.11g 54Mbps (optional |